SPECIFICATIONS
– Discrete, Random access, fully automated, STAT Sample Priority
– 240 tests per hour, 420 tests per hour with ISE
– 40 Reagent Positions, 40 Sample Positions in cooling compartment (4-12 °C)
– End Point, Fixed Time, Kinetic
– Colorimetric / Turbidimetry
– Linear/non-linear multi-point calibration
– Automatic Probe cleaning, liquid level detection, collision protection,Carry – over < 0.1%
– 8 wavelengths: 340 – 670 nm
– Automatic pre-and post-dilution for abnormal sample Dilution Rate 1:200
– Reaction Volume from 150 – 500 μl
– Automatic cuvette wash & cell blank check
– Real-time reagent levels monitoring, displaying and alerting function
– Stand Alone mixing bar
– CE-marked applications with Diasystem reagents for Routine Chemistry, Immunoturbidimetry, Drugs of Abuse & TDM