Dimensions
height: 56 in (142.24 cm)
width (wheels in outmost position): 28.5 in (72.39 cm)
depth (wheels in outmost position): 29.5 in (74.93 cm)
weight: 162.7 lbs (73.8 kg)
battery
operating time: a new, fully charged 1-hour bps can operate the ventilator for at least one hour. This 1-hour bps will not run the compressor or a humidifier. It is recommended that the 1-hour battery pack be replaced every two years.
Recharge time: 20-hours while connected to ac power
gas mixing system
range of flow from the mixing system: can be set to 150 l/min standard temperature and pressure, dry (stpd). Additional flow is available (up to 30 l/min for neonatal circuit type, up to 80 l/min for pediatric circuit type, and up to 200 l/min for adult circuit type) for compliance compensation.
Leakage from one gas system to another: meets standard
operating pressure range: 35 to 100 psi (241 to 690 kpa)
air/oxygen regulator bleed: up to 3 l/min
oxygen and air inlet supplies (not included)